C-195
P/N 156452
REVISION W
12/2007
STACKABLE LCC/LCD OVEN
INSTRUCTION MANUAL
Model Atmosphere
Volts Amps
Hz Heater
Watts
Phase
LCC/D1-16-3 Air 240 14.8 50/60 3,000 1
LCC/D1-16N-3
Nitrogen 240 14.0 50/60 3,000 1
LCC/D1-51-3 Air 240 27.7 50/60 6,000 1
LCC/D1-51N-3
Nitrogen 240 27.7 50/60 6,000 1
Model numbers may include a “V” for silicone free construction.
Model numbers may begin with the designation LL*1-*, indicating
Models without HEPA filter.
Prepared by:
Despatch Industries
8860 207
th
St. West
Lakeville, MN 55044
Customer Service 800-473-7373
Seitenansicht 0
1 2 3 4 5 6 ... 104 105

Inhaltsverzeichnis

Seite 1

C-195 P/N 156452 REVISION W 12/2007 STACKABLE LCC/LCD OVEN INSTRUCTION MANUAL Model Atmosphere Volts Amps Hz Heater Watts Phase LCC/D1-

Seite 2

1 INTRODUCTION The Stackable LCC/LCD series OVEN offers HEPA filtration for processes where minimization of contamination is essential. The removab

Seite 3

91 LCC1-51N-3 Drawing 163141*01

Seite 4

92 LCC1-51N-3 Drawing 163141*02

Seite 5

93 LCC1-51N-3 Drawing 163141*03

Seite 6

94 LCC1-51N-3 Drawing 163141*04

Seite 7

95 APPENDIX B: Temperature Scale Conversion and Optional MRC5000 Setup Temperature Scale Conversion (C/F) The Protocol Plus controller can be ope

Seite 8 - TABLE OF CONTENTS

96 Optional MRC5000 Recorder Setup The temperature is retransmitted to the Recorder from the Controller. Set up the Recorder as follows: 1. Mak

Seite 9

2 Features • Despatch Protocol Plus microprocessor-based digital programmable control, with simultaneous digital readout of both setpoint and actu

Seite 10 - INTRODUCTION

3 SPECIFICATIONS Model numbers that begin with the designation LL*1-* are without HEPA filter, and have the same physical and power characteristics

Seite 11 - Options

4 Capacities Model LCC/D1-16-3 LCC/D1-16N-3 LCC/D1-51-3 LCC/D1-51N-3 Maximum Load (Lbs) 200 200 200 200 Maximum Shelf Load (Lbs) 50 50 25

Seite 12 - SPECIFICATIONS

5 Temperature Model LCC/D1-16 LCC/D1-16N LCC/D1-51 LCC/D1-51N 40°C – 100°C 7 7 5 5 40°C – 200°C 30 30 27 27 40°C – 260°C 45 45 35 35

Seite 13 - Capacities

6 INSTRUCTIONS The INSTRUCTIONS section provides directions on unpacking, installation, operation and maintenance of the Despatch Stackable LCC Ser

Seite 14 - Temperature

7 LCC/LCD/LLC/LLD DOOR LOCK MANUAL OVERIDE Requires ¼” multi-tip driver handle. One tip included per oven. Note recess in end of tip. Separate,

Seite 15 - INSTRUCTIONS

8 Set-up 1. Select the location for installing your oven. Single Oven Placement Place the oven on a bench top or other framework capable of hold

Seite 16

9 WARNING: Never allow drain to be plugged as a hot oven will generate a small amount of steam when the water is first turned on. Steam bu

Seite 17

10 LCC/LCD1-16 & 51-3 WITH WATER COOLING UTILITY CONNECTIONS Nitrogen or Clean Dry Air Inlet. 70 to 80 PSI (4.83 to 5.52 Bar). Used to pu

Seite 18

ii Dear Customer, Thank you for choosing Despatch Industries. We appreciate the opportunity to work with you and to meet your heat processin

Seite 19

11 LCC/LCD1-16N & 51N-3 UTILITY CONNECTIONS Nitrogen or Clean Dry Air Inlet. 70 to 80 PSI (4.83 to 5.52 Bar). Used for Nitrogen/Clean Dry

Seite 20

12 LCC/LCD/LLC/LLD1-16-3 AND LCC/LCD/LLC/LLD1-51-3 EXHAUST (ALL SERIES) Exhaust located on left side of oven, towards the rear. Exhaust Stack Mus

Seite 21 - (ALL SERIES)

13 STACKABLE LCC-LCD-LLC POWER CONNECTION Ground Buss Panduit to protect line wire between conduit and disconnect switch. Power connection a

Seite 22

14 ½” conduit for customer wiring from rear compartment REAR COMPARTMENT – PANEL REMOVED Slot in Chassis FRONT SWING PAN

Seite 23

15 Equipment nameplate REAR PANEL Conduit slot

Seite 24

16 Wiring NOTE that the oven must be directly hardwired to the disconnect switch on the equipment panel. Please see the picture for the wiring cond

Seite 25

17 NOTE: Repairing the damaged filter unit, particularly the medium, should not be attempted by the user. Any unit so repaired must be retested to

Seite 26 - HEPA Filter Installation

18 • Check the corner joints of the frame for adhesive sealing and tightness. • Check that the gaskets are cemented firmly to the filter frame

Seite 27 - HEPA Filter Burn-off

19 is used at temperatures above 180°C / 356°F. There will be smoke, possibly a pungent odor, and a light residue on interior surfaces. This is t

Seite 28

20 NOTE: If it is necessary to move the equipment after the burn-off process, considerable care should be used. The binder which gives strength t

Seite 30 - OVEN OPERATION

21 OVEN OPERATION Oven The stackable LCC oven is a class 100 clean room oven with HEPA (High Efficiency Particulate Air) filtration. This oven is

Seite 31 - System Control

22 System Control A Despatch Protocol Plus controller/high-limit device is used to control the unit. This is located on the control panel of the o

Seite 32

23 Upper Portion of Control Panel (optional beacon light installed) Lower Portion of Control Panel Manual Unlock: In the event of power fa

Seite 33 - HEPA Filters

24 HEPA Filters HEPA (High Efficiency Particulate Air) filters are used to limit particulate size in the work chamber to 0.3 microns or less. NO

Seite 34 - Class 100 Testing

25 Handling The filter is shipped in the original carton or package that the filter manufacturer uses. This will give good storage and maximum pro

Seite 35 - Filter Unit Replacement

26 Validation Testing Based on the issues discussed in this section, Despatch recommends the following test sequence for pharmaceutical Class 100 o

Seite 36

27 • Water droplets in airstream through filter, free water (RH = 100%), will saturate filter very quickly and may cause burnout or holes in burned

Seite 37

28 HEPA Filter / Magnehelic Pressure Gauge The LCC Series oven is equipped with a Magnehelic pressure gauge which measures the pressure in front of

Seite 38 - OPERATING

29 OPERATING Users and operators of this oven must comply with operating procedures and training of operating personnel as required by the Occupati

Seite 39 - Pre-Startup Checklist

30 Pre-Startup Checklist • Know the system. Read this manual carefully. Make use of its instructions and explanations. The know how of safe, co

Seite 41

31 Operating Procedure Starting the Oven 1. Turn the yellow/red DISCONNECT SWITCH to ON. 2. Press the POWER switch to ON. • The DOOR REL

Seite 42

32 • Press the Select key until Profile is displayed (note you can press the Run key at any time to activate Profile Mode). • Press the key to d

Seite 43

33 meter to the desired setting (see table below for suggested values). This program energizes the purge solenoid valve. 7. The second segment is

Seite 44 - Maintenance

34 LCC1-16N-3 AND LCC1-51N-3 N2 NEEDLE VALVE Nitrogen MAINTAIN needle valve. Turn Event 3 (Purge) to ON, this is the “Purge” cycle. Adjust nitrog

Seite 45 - CONTROL INSTRUCTIONS

35 Maintenance Do not attempt any service on this oven before opening the main power disconnect switch. Checklist • Keep equipment clean. Grad

Seite 46

36 CONTROL INSTRUCTIONS The special features of the Protocol PlusTM control include: • PID tuning • Ramp/Soak programming of up to 64 segments •

Seite 47 - Fast Start Mode

37 Protocol Plus Faceplate and Wiring Diagram

Seite 48 - Indicators

38 Operating Modes The Protocol Plus control has five modes of operation available: Stopped Mode: All control and relay outputs are off. Stopped M

Seite 49 - Key Functions

39 High Limit The control has an integrated high limit function which will disable the heater output when tripped. If the high limit does trip, the

Seite 50 - Outputs

40 Displays The Protocol Plus control has two displays. A dedicated LED upper display shows the oven temperature. A two-line LCD lower display provi

Seite 52

41 Outputs The Protocol Plus control has seven different outputs available. • Heating output: The control output is a DC voltage open-collector ou

Seite 53 - Operation

42 Relay (Continued) Use the Relay Card Optional Ay p/n 144562 to add relays to the standard controller. Each relay card contains two relays (maximum

Seite 54 - Timer Mode

43 INSTRUCTIONS Start-Up These instructions are provided as a quick reference for operating the Protocol Plus control. If the Profile Mode is to be

Seite 55 - Auto Start Mode

44 Operation Manual Mode Press the Select key until Manual is displayed (note you can press the Run key at any time to activate Manual Mode). 1.

Seite 56

45 Timer Mode 1. Press the Select key until Timer is displayed (note you can press the Run key at any time to activate Timer Mode). 2. Press the

Seite 57

46 Profile Mode 1. Press the Select key until Profile is displayed. “None” may be displayed if a profile has not been selected or no profiles ente

Seite 58

47 3. On Auto Start page Enable set to No, will deactivate Auto Start Mode. Note that once you activate Auto Start, you can continue to use all op

Seite 59

48 Instructions for Setup Mode Pages Program Page (Defaults on Page 66) Programming of the profiles is provided on the Program Page. Eight profile

Seite 60

49 Menu Item Display Description Ramp Time Seg 1 Pro-1 Seg-1 Ramp Time Ramp time for segment 1 of profile Event 1 Set Value* Pro-1 Seg-1 Ramp E

Seite 61

50 Profile # There are eight profiles available. Segment# Recipe segments 1 through 8 may be programmed, each with its own set of events, ramp a

Seite 62

vi PREFACE This manual is your guide to the Despatch stackable LCC ovens. It is organized to give you the information you need quickly and easi

Seite 63

51 Sample Profile (Blank Program Table on Page 67) Programming Table Profile Number____1______ Profile Name__________ Ramp Soak Events Events

Seite 64

52 Auto Start Page (Optional, Defaults on Page 68) If the optional real time clock has been installed, the Auto Start Page can be configured to au

Seite 65

53 Menu Item Display Description Range Enable Autostart Auto Start Enable Enable (yes) or disable (no) the Autostart function No, Yes Sunday mod

Seite 66

54 PID Page (Defaults on Page 68) The PID Page contains parameters which control the response to the setpoint and process variable input. To acce

Seite 67

55 Control Page (Defaults on Page 69) The Control Page contains various parameters which control miscellaneous functions. To access the Control Pa

Seite 68

56 Communication Page (Optional, Defaults on Page 69) The Communication Page contains parameters for the communications feature. To access the Com

Seite 69

57 Relay Outputs Page (Optional, Defaults on Page 70) The Relay Outputs Page configures the four alarm/event outputs. Each output has a dedicated

Seite 70

58 Test Page (Defaults on Page 71) The Test Page contains parameters which allow manual control of the heat control and optional relay outputs and

Seite 71

59 Zone Calibration Page (Defaults on Page 71) The Zone Calibration Page allows adjustment of the displayed temperature versus the actual temperat

Seite 72

60 As an example, the control is displaying 400°F with the setpoint being 400°F, but the center of the oven chamber is actually 395°F. This can o

Seite 73 - Digital Inputs (optional)

vii Revision B (6-01): Various corrections Revision C (11-01): Corrections, addition of schematic drawings Revision D (1-02): Corrections, upd

Seite 74 - Error Messages and Alarms

61 Sensor Calibration Page (Defaults on Page 71) The Sensor Calibration Page has parameters which can change the internal calibration of the tempe

Seite 75

62 displayed. 19. Connect a calibration instrument with a type J thermocouple output to the control sensor input. 20. Connect a voltage measureme

Seite 76

63 Enable Page (Defaults on Page 72) The Enable Page controls access to the other Setup Pages. The setpoint minimum and maximum values, and the se

Seite 77

64 Digital Inputs (optional) The Protocol Plus control can be run by external inputs wired to the control from an external source such as a PLC or

Seite 78

65 Error Messages and Alarms The Alarm Status Hi-limit LED is flashing. This indicates a problem with the thermocouple, or the Hi-limit setpoint h

Seite 79

66 Quick Reference and Default Values Program Page Menu Item Display Default Range Setting Ramp Time Seg 1 Pro-1 Seg-1 Ramp Time 00:00 00m00

Seite 80

67 Programming Table Profile Number__________ Profile Name__________ Ramp Soak Events Events Seg-ment Time 1 2 3 4 Temp-erature Time 1 2 3 4 1

Seite 81

68 Autostart Menu Item Display Default Range Setting Enable Autostart Auto Start Enable No No, Yes Sunday mode Auto Start Sun Mode Off Off, Manu

Seite 82 - Technical Specifications

69 Control Menu Item Display Default Range Setting Cycle Time Control Cycle Time 1 1 to 60 seconds High limit setpoint Control Hi-Lim SP Max Hi

Seite 83 - APPENDIX A: DRAWINGS

70 Real Time Clock Menu Item Display Default Range Setting Day of the week Clock Day Mon Sun, Mon, Tue, Wed, Thu, Fri, Sat Time of day Clock HH:

Seite 84 - Drawing 150008*01

viii TABLE OF CONTENTS INTRODUCTION... 1

Seite 85 - Drawing 150008*02

71 Test Menu Item Display Default Range Setting Heater output Test HeatOut Off On High limit relay Test HiLimOut Off On Relay 1 output Test Rl

Seite 86 - Drawing 150008*03

72 Enable Page Menu Item Display Default Range Setting Profiles Enable Profile 1-8 Yes Yes or No Autostart Enable Auto Start No Yes or No PID

Seite 87 - Drawing 150047*01

73 Technical Specifications UL, cUL listed: UL file E136675 CE compliance to: • EMC Directive 89-366/EEC as modified by 93/68/EEC • European Sta

Seite 88 - Drawing 150047*02

74 APPENDIX A: DRAWINGS The following pages contain electrical drawings for the LCC1-16, LCC1-16N-3, LCC1-51-3 and LCC1-51N-3 ovens.

Seite 89 - Drawing 150047*03

75 LCC1-16-3 Drawing 150008*01

Seite 90 - Drawing 150044*01

76 LCC1-16-3 Drawing 150008*02

Seite 91 - Drawing 150044*02

77 LCC1-16-3 Drawing 150008*03

Seite 92 - Drawing 150044*03

78 LCC1-16N-3 Drawing 150047*01

Seite 93 - Drawing 159135*01

79 LCC1-16N-3 Drawing 150047*02

Seite 94 - Drawing 159135*02

80 LCC1-16N-3 Drawing 150047*03

Seite 95 - Drawing 159135*03

ix Communication... 42 Optional Software...

Seite 96 - Drawing 159135*04

81 LCC1-51-3 Drawing 150044*01

Seite 97 - Drawing 159159*01

82 LCC1-51-3 Drawing 150044*02

Seite 98 - Drawing 159159*02

83 LCC1-51-3 Drawing 150044*03

Seite 99 - Drawing 159159*03

84 LCC1-51N-3 Drawing 159135*01

Seite 100 - Drawing 163141*01

85 LCC1-51N-3 Drawing 159135*02

Seite 101 - Drawing 163141*02

86 LCC1-51N-3 Drawing 159135*03

Seite 102 - Drawing 163141*03

87 LCC1-51N-3 Drawing 159135*04

Seite 103 - Drawing 163141*04

88 LCC1-51-3 Drawing 159159*01

Seite 104

89 LCC1-51-3 Drawing 159159*02

Seite 105

90 LCC1-51-3 Drawing 159159*03